Operators’ Manual Monroe Electronics, Inc. Model 288B Charge Plate Monitor P/N 0340174 288-1/100 V2.10 022713 10
10 SETUP - OPTION Menu: Options are for display backlighting, power off, and beep settings. The first two options will apply when the 288 unit is
11 Section 4 OPERATION MANUAL OPERATION: From the manual screen the user can choose to run a positive decay, negative decay or balance. Also the
12 Figure 10. +Decay Test Result: REVIEW: review the test results. MANUAL: go back to MANUAL screen. +DECAY: start a +Decay test. REVIEW: di
13 of the display will flash, which signifies that the average voltage computation has overflowed and is no longer updating based on new data. The
14 AUTOMATIC OPERATION: Once the desired test parameters are set up, the user simply selects the “start” button to begin the tests. All the parame
15 When the auto sequence tests are finished, data will be shown as follows: Figure 16. Auto Sequence Test Review: NxtCyc: highlight the next d
16 From the MAIN Screen (Figure 1), press SETUP – AUTO. Then highlight the A.Setup 1, press DISPLAY. The following screen will appear: Figure 17. S
17 ADDITIONAL FEATURES: Group and Location – By accessing this screen the tests can be organized to reflect the ionizers’ locations. There are up
18 Figure 19. Review Data: From MAIN screen (Figure 1), press DATA RvwMan: Review Manual test data. See Figure 11. RvwAuto: Review Auto sequence da
19 Section 5 MAINTENANCE Precautions — User maintenance should normally be limited to keeping the instrument clean and free from physical damage.
2 CONTENTS Specifications Section 1 Specifications Page 4 Accessories included Page 5 What this instrument does Page 6 Physical description Pag
20 Section 6 REFERENCES Documents associated with ionization: ESD Association Standard — ANSI/ESD STM3.1 –Ionization ESD Association Advisory – E
21 WARRANTY Monroe Electronics, Inc., warrants to the Owners, this instrument to be free from defects in material and workmanship for a period of
22 RETURN POLICIES AND PROCEDURES FACTORY REPAIR Return authorization is required for factory repair work. Material being returned to the factory
3
4 Section 1 Specifications All specifications are referred to plate voltage unless otherwise specified. Display 240 x 64 character/graphic
5 Charge Plate Capacitance 20 pF ±10% Zero Drift < 100 mV /sec (no incident ion flow) Self Discharge < 200 mV/sec Pea
6 What this instrument does — This instrument is a charged-plate monitor for evaluating the performance of ionization systems. As such, it perform
7 Section 2 GENERAL INFORMATION Model 288 performs manual or automatic decay and balance tests for qualification and periodic verification of ioni
8 Section 3 SETUP First things first Upon power up, you are presented momentarily with an identification screen, which includes the software revis
9 RAM/EE – The first screen under this option asks, “Clear all test data – Y/N?” A “Y(es)” response will permanently erase all accumulated test da